OPTITEMP TCA-M40
Thermocouple (TC) cable sensor for machinery and high temperature applications
- Compression fitting connection
- Mineral insulated, with connection head type BA, available with various sheath materials
- -40…+1250°C / +2282°F
OPTITEMP TCA-M40
Thermocouple (TC) cable sensor for machinery and high temperature applications
- Compression fitting connection
- Mineral insulated, with connection head type BA, available with various sheath materials
- -40…+1250°C / +2282°F
Overview
The OPTITEMP TCA-M40 is a mineral insulated (MI) thermocouple with a connection head type BA. The MI cable sensor is particularly suitable for applications where the temperature of the environment exceeds the maximum allowable temperature of standard cable material (e.g. in test bench applications). The MI cable can be easily formed to fit into limited spaces. Its small diameter makes it perfect to provide fast response times.
The OPTITEMP TCA-M40 is mechanically and thermally very stable. It is available with thermocouple type J, K or N, single (standard) or double (optional), according to IEC 60584 class 1. The connecting wires are colour coded according to IEC 60584. The measuring point can be ordered as isolated (standard) or grounded (optional). The MI thermocouple sensor is best used for measuring high process temperatures of gases and liquids (up to +1250°C / +2282°F), at low to medium pressures and flow velocities.
Product features and options depend on device configuration: Contact us to make sure that your preferred functionality can be combined in one device.
- High temperature measurement in limited spaces
- Flexible and hand formable MI cable sensor
- Mechanically and thermally stable
- Designed for fast response times
- Connection head BA (IP65) available with optional head-mounted transmitter
- Sheath material: 1.4841 / AISI 310, 2.4816 / Alloy 600 or Pyrosil®
- Suitable for installation where a connection terminal already exists
- Tolerance class 1 according to IEC 60584-2 as standard
- Measuring point isolated or grounded